What is Software as a Medical Device?
The International Medical Device Regulators Forum (IMDRF) defines SaMD as software intended to be used for one or more medical purposes that performs those purposes without being part of a hardware medical device. FDA uses this framing in its digital health materials. The key idea is that the software itself is the device.
Two related terms help. Software in a Medical Device (SiMD) is software that is part of a hardware device, such as an infusion pump's control firmware. Device software functions is the broader FDA term for software functions that meet the statutory definition of a device, whether standalone or embedded. Non-device software functions, such as billing or general administrative tools, are outside this scope.
The legal hook is the definition of "device" in section 201(h) of the Federal Food, Drug, and Cosmetic Act, which turns on intended use, including claims made in labeling, marketing and sales materials. How you describe the product matters as much as what the code does.
02
Is your software a device? A decision table
Use this table as a screening aid. It cannot replace a formal assessment, and several rows depend on details FDA examines closely.
| Software function (fictional example) | Typical posture | Why it matters |
|---|---|---|
| Clinic scheduling and billing app | Not a device | No medical purpose; outside device definition |
| Fitness app encouraging activity and sleep habits, with low-risk wellness claims | Likely general wellness, generally not actively regulated | Intended for general wellness without reference to diseases or, in some cases, with claims tied to a well-accepted benefit |
| Medication reminder with no dosing or interpretation | Often enforcement discretion or non-device, depending on claims | FDA's policy for device software functions lists categories where it does not intend to enforce requirements |
| Tool displaying patient information and guideline-based recommendations to clinicians who can review the basis | May be non-device CDS if all statutory criteria are met | Section 520(o)(1)(E) and FDA's 2022 CDS guidance |
| Software that analyzes a patient's image, signal or pattern to flag a condition | Likely a device | Analyzing medical images or signals is outside the CDS exclusion |
| App calculating insulin doses from glucose readings | Likely a device | Drives treatment decisions with direct patient risk |
| AI model giving time-critical alerts about deterioration | Likely a device | Time-critical, specific output; risk to patients if wrong |
03
Clinical decision support: device or non-device?
The 21st Century Cures Act added section 520(o)(1)(E) to the FD&C Act, which excludes certain CDS software functions from the device definition. FDA's final guidance on Clinical Decision Support Software (September 2022) interprets four criteria. In summary, a function may be non-device CDS only if it:
- Is not intended to acquire, process or analyze a medical image, a signal from an in vitro diagnostic device, or a pattern or signal from a signal acquisition system.
- Is intended to display, analyze or print medical information about a patient or other medical information, such as peer-reviewed clinical studies or clinical practice guidelines.
- Is intended to support or provide recommendations to a health care professional about prevention, diagnosis or treatment of a disease or condition.
- Is intended to enable the health care professional to independently review the basis for the recommendations, so that the professional does not rely primarily on them to make a clinical decision.
04
General wellness and enforcement discretion
FDA's General Wellness policy describes low-risk products intended only for general wellness, such as encouraging healthy lifestyle choices, that FDA does not intend to regulate as devices. Claims that refer to diagnosing, treating or preventing a specific disease usually take a product out of that category.
Separately, FDA's policy for device software functions and mobile medical applications identifies functions that may meet the device definition but pose lower risk, so FDA exercises enforcement discretion and does not currently expect premarket submissions. Enforcement discretion is a statement about FDA's priorities, not an exemption in law, and it can change. Examples in the policy include some tools that help users organize or track health information, or provide simple educational content, without interpreting data to drive diagnosis or treatment.
Practical advice: write your intended-use statement first, review every marketing claim against it, and re-review whenever you add a feature such as risk scores, alerts or dosing suggestions. Feature creep is the most common way a wellness app becomes a device.
05
FDA premarket pathways: 510(k), De Novo and PMA
If your software is a device, its classification determines the pathway. Devices are classified by risk into Class I, II or III. Many software devices land in Class II, though classification depends on the specific product and the available predicates.
FDA's Q-Submission program lets sponsors ask for feedback on classification, testing and study plans before a formal submission. It is often the cheapest way to reduce uncertainty. For software devices, FDA's guidance on premarket submission content for device software functions describes the documentation it expects, scaled to the software's level of concern.
| Pathway | When it fits | Standard | Planning notes |
|---|---|---|---|
| Exempt or enforcement discretion | Low-risk functions that FDA exempts or chooses not to enforce | No premarket submission, but other general controls may still apply | Confirm status in writing with counsel; consider a Q-Submission to FDA |
| 510(k) | Moderate-risk devices with a legally marketed predicate | Substantial equivalence to a predicate | Needs predicate research, performance testing, software documentation and cybersecurity documentation |
| De Novo | Novel low-to-moderate risk devices with no suitable predicate | Reasonable assurance of safety and effectiveness with general and special controls | Creates a new classification that later products can use as a predicate |
| PMA | High-risk Class III devices | Valid scientific evidence of safety and effectiveness | Typically requires substantial clinical evidence and a longer, costlier review |
06
Quality system, IEC 62304 and ISO 14971
Device manufacturers must operate under FDA's quality system requirements, which FDA has been aligning with ISO 13485 through the Quality Management System Regulation (QMSR). Check the current effective dates and requirements with your regulatory advisor. Design controls require you to plan, verify and validate design, keep a design history file and manage changes.
Two standards dominate software work. IEC 62304 defines the medical device software life cycle: planning, requirements, architecture, detailed design, implementation, verification, release and maintenance, with software safety classes A, B and C reflecting potential harm. ISO 14971 defines risk management: hazard identification, risk estimation, evaluation, control, and residual risk review across the product lifecycle. FDA recognizes both as consensus standards.
In practice this changes how engineering works. Requirements are traceable to tests, third-party components are tracked as SOUP (software of unknown provenance), changes go through documented impact analysis, and risk controls become tested features. Teams that bolt this on afterward find it costly, so decide early whether a function needs it. Our medical device software development practice builds with IEC 62304-aware processes.
07
Cybersecurity (section 524B) and predetermined change control plans
Section 524B of the FD&C Act, added in 2022, requires premarket submissions for "cyber devices" to include a plan to monitor and address postmarket vulnerabilities, processes that provide reasonable assurance of cybersecurity, a software bill of materials, and other information FDA requires. A cyber device generally includes software, can connect to the internet, and has characteristics that could be vulnerable to cybersecurity threats. FDA's premarket cybersecurity guidance explains expectations in detail, including threat modeling and secure product development practices.
For AI-enabled software, FDA issued final guidance in December 2024 on predetermined change control plans (PCCPs). A PCCP lets a manufacturer describe, in advance, certain planned modifications, the methods to develop, validate and implement them, and an assessment of their impact, so that those changes can be made without a new submission if FDA authorizes the plan with the device. Whether a PCCP is appropriate and sufficient depends on the product and FDA's review.
- Keep an accurate SBOM from the first sprint; rebuilding one later is painful.
- Plan vulnerability monitoring and coordinated disclosure processes before launch.
- For learning or frequently updated models, define in advance what can change, how you will test it and how you will monitor performance.
- Do not assume a PCCP is available or granted. It is part of a marketing submission and subject to FDA review.
08
How FDA fits with HIPAA, state law and other markets
FDA regulation of a device does not replace privacy law. A SaMD sold to providers may also make you a HIPAA business associate, while a consumer SaMD may fall under the FTC Health Breach Notification Rule and state consumer health data laws. See our HIPAA guide for startups for how those fit together.
If you plan to serve EU users, the EU Medical Device Regulation and, where AI is used, the EU AI Act may apply in addition to GDPR, which treats health data as a special category. Get market-specific advice before launching abroad.
09
A practical path for product teams
We help teams scope and build regulated and non-regulated software through HealthTech product development, medical device companion app development, digital therapeutics development and product consulting. We do not provide legal or regulatory advice, and we cannot guarantee any FDA outcome.
- Write a precise intended-use and indications-for-use statement.
- Screen each feature against the device definition, CDS criteria and enforcement discretion policies.
- Engage regulatory counsel or a regulatory consultant to confirm classification; consider a Q-Submission to FDA for uncertain cases.
- If device functions exist, set up design controls, risk management and a software life cycle process before building them.
- Separate device functions from non-device functions in your architecture where it reduces regulatory scope.
- Plan cybersecurity documentation, SBOM and postmarket monitoring from the start.
Frequently asked questions
What does SaMD stand for and who defines it?
SaMD means Software as a Medical Device. The International Medical Device Regulators Forum defines it as software intended for one or more medical purposes that performs them without being part of a hardware medical device. FDA uses the concept in its digital health policies but regulates through the FD&C Act's device definition.
Is clinical decision support software always regulated by FDA?
No. Section 520(o)(1)(E) excludes certain CDS functions from the device definition if all criteria are met, including that a health care professional can independently review the basis for recommendations. Software analyzing images or signals, or giving time-critical directives, is more likely to be a device.
How long does FDA clearance take?
It varies widely by pathway, product risk, quality of evidence and FDA questions. Preparing the quality system, testing and documentation often takes longer than the review itself. Treat any timeline as a planning estimate, and ask regulatory counsel or a consultant for a view specific to your product.
Do AI features need a predetermined change control plan?
Not necessarily. A PCCP is optional and applies to AI-enabled device software functions that are authorized through a marketing submission. FDA's December 2024 final guidance describes what to include. Whether it suits your product depends on how the model will change and on FDA's review.
Does a general wellness app need FDA clearance?
Generally not, if it is intended only for general wellness and makes no claims about diagnosing, treating or preventing a disease. Claims, features and marketing can change that status. Re-check each time you add risk scores, alerts or disease-specific language.
Sources and further reading
- FDA: Software as a Medical Device (SaMD)
- FDA: Clinical Decision Support Software guidance
- FDA: General Wellness: Policy for Low Risk Devices
- FDA: PCCP for AI-Enabled Device Software Functions
- FDA: Cybersecurity in Medical Devices
This page is general information, not legal or regulatory advice. Requirements depend on your product, data and organization; consult qualified counsel for your situation.